Filters and extension lines
Endoxin Filters
Beskrivning
Code 807.204: Filter-A-line Neonatal/Paediatric 0.22micrometer I.V. filter 96-hour filter for the removal of inadvertent bacterial contaminants,… Läs mer
Indikation
Extension tubes and endotoxin filters for the management of intravenous lines for premature infants, full-term newborns and children.
Beskrivning
Code 807.204: Filter-A-lineNeonatal/Paediatric 0.22micrometer I.V. filter
96-hour filter for the removal of inadvertent bacterial contaminants, associated endotoxins, particulates, and air - Male and female Luer-lock fittings - Not for use with lipids, blood or blood products.
Priming volume: 0.43ml - Flow rate: 5.80ml/min
Code 807.205: Filter-A-line
Neonatal/Paediatric 0.22micrometer I.V. filter with extension line
96-hour filter for the removal of inadvertent bacterial contaminants, associated endotoxins, particulates and air.With microbare tubing and clamp. Male and female Luer-lock fittings. Not for use with lipids, blood or blood products.
Priming volume: 0.59ml - Flow rate: 2.5ml/min – Length: 24cm
Other endotoxin filters are available under the codes 836.201, .202, .204, .205 and .210.
Code 836.201: Ready-set "feed"
Polyurethane double-lumen extension tube
Lumen 1 (15cm long) incorporates a clamp, a 0.22micrometer endotoxin, bacterial, particulate, and air filter, 96 hours with low protein adsorption membrane. This air-eliminating filter is not suitable for blood, blood products or lipids. Priming volume: 0.84ml - Flow rate: 3.5ml/min
The non-return valve on the distal part of the line avoids the occlusion of the filter by reflux of lipidic solutions if the other line is used for that purpose.
Lumen 2 (2cm long) incorporates a closed-system needleless connector (Bionector) and can be used either for intermittent injections of medication or for continuous infusions (e.g. lipidic solutions). Priming volume: 0.24ml - Flow rate: 75ml/min
Total priming volume: 0.90ml
Code 836.202: Ready-set "complete"
Polyurethane extension tube with 2 access ports incorporating non-return valves, clamps and a 0.22micrometer endotoxin, bacterial, particulate, and air filter, 96 hours with low protein adsorption membrane. This air-eliminating filter is not suitable for blood, blood products or lipids.
Access ports: Priming volume: 0.95ml (x2) - Flow rate: 2.5ml/min (x2)
Length: 20.5cm - Total priming volume: 1.1ml - Flow rate: 2.5ml/min
Code 836.204: Safe-t-piece
Polyurethane extension tube with clamp and 0.22micrometer endotoxin, bacterial, particulate, and air filter, 96 hours with low protein membrane adsorption. This air-eliminating filter is not suitable for blood, blood products or lipids.
The distal end is a "T" piece with a male Luer-lock hub on one side, a latex-free injection membrane on the other side.
Proximal end: closed-system needleless connector (Bionector).
Length: 11cm - Flow rate: 4.5ml/min - Priming volume: 0.64ml
This product is mainly aimed at neonatal/paediatric units where connections are made directly to the cannula.
Code 836.205: Ready-set "unite"
Polyurethane extension tube with 3 access ports.
Main line (20.5cm long): 2 access ports incorporating non-return valves and clamps. This line is also fitted with a 0.22micrometer endotoxin, bacterial, particulate, and air filter, 96 hours with low protein adsorption membrane. This air-eliminating filter is not suitable for blood, blood products or lipids. Priming volume: 0.98ml – Flow rate: 2.5ml/min
The non-return valve on the distal part of the line avoids the occlusion of the filter by reflux of lipidic solutions if the other line is used for that purpose.
Lumen 2 (2cm long): closed-system needleless connector (Bionector) and clamp, can be used either for intermittent injections of medication or for continuous infusions (e.g. lipidic solutions). Priming volume: 0.26ml - Flow rate: 75ml/min
Total priming volume: 1.3ml
Code 836.210: Ready-set
Polyurethane extension tube with 3 access ports. Total priming volume: 1.4ml
Main line (20.5cm long): 2 access ports incorporating closed-system needleless connectors (Bionector), non-return valves and clamps. This line is also fitted with a 0.22micrometer endotoxin, bacterial, particulate, and air filter, 96 hours with low protein adsorption membrane. This air-eliminating filter is not suitable for blood, blood products or lipids. Flow rate: 2.5ml/min (x2) – Priming volume: 1ml (x2)
The non-return valve on the distal part of the line avoids the occlusion of the filter by reflux of lipidic solutions if the other line is used for that purpose.
Lumen 2 (2cm long): closed-system needleless connector (Bionector) and clamp, can be used either for intermittent injections of medication or for continuous infusions (e.g. lipidic solutions). Priming volume: 0.26ml - Flow rate: 75ml/min
Referenser och funktioner
Extension tube and/or filter | Filter | Extension tube | Förpackning | ||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Favourites | Kod | Priming volume ml | Flow rate ml/min | Maximum pressure in use bar | Length cm | - µm | Int. Ø mm | Ext. Ø mm | Length cm | Priming vol. lumen 1 ml | Priming vol. lumen 2 ml | Total priming volume ml | Flow rate lumen 1 ml/min | Flow rate lumen 2 ml/min | Pressure tested at bar | Antal/Avd.förpackning | Antal/Trp.förpackning |
807.204 | 0.43 | 5.80 | 2 | - | - | - | - | - | - | - | - | - | - | - | 10 | 240 | |
807.205 | 0.59 | 2.5 | 2 | 24 | - | - | - | - | - | - | - | - | - | - | 10 | 120 | |
836.201 | - | - | - | - | 0.22 | 1.5 | 2.5 | 2 - 15 | 0.24 | 0.84 | 0.90 | 75 | 2.5 | 2 | 10 | 120 | |
836.202 | - | - | - | - | 0.22 | 1.5 | 2.5 | 20.5 (x2) | 0.95 | 0.95 | 1.1 | 2.5 | 2.5 | 2 | 10 | 120 | |
836.204 | - | - | - | - | 0.22 | 1.5 | 2.5 | 11 | - | - | 0.64 | 4.5 | - | 2 | 10 | 120 | |
836.205 | - | - | - | - | 0.22 | 1.5 | 2.5 | 2 - 20.5 (x2) | 0.98 (x2) | 0.26 | 1.3 | 2.5 | 75 | 2 | 10 | 120 | |
836.210 | - | - | - | - | 0.22 | 1.5 (x3) | 2.5 (x3) | 2 - 20.5 (x2) | 1.0 (x2) | 0.26 | 1.4 | 2.5 (x2) | 75 | 2 | 10 | 120 |
Ytterligare information
-
Innehåller latex Nej
-
Innehåller animaliska produkter Nej
-
Pyrogenfri Ja